medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge- BEQ-TOP 37200 E-Circuit Bioline Heparin

Recalled Product

Maquet Getinge- BEQ-TOP 37200 E-Circuit Bioline Heparin Material: 701055388R01


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000076762 3000077899 3000078013 3000084649