medical Class II Updated 2021-05-26

Bard Access Systems Inc. recalls REF 7617408D, Groshong NXT ClearVue Catheter, Maximal Barrie

Recalled Product

REF 7617408D, Groshong NXT ClearVue Catheter, Maximal Barrier Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035262 - Product Usage: is designed for use when central venous catheterization is prescribed.


Hazard / Issue

Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot No. reev2091, reex1265, reew3144, redx1464, reen4697, REEP0632