medical MODERATE Updated 2026-04-08

Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00

See all recalls from Reflexion Medical, Inc. →

Recalled Product

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

Issued by

FDA

Affected States: CA, TX, PA, CT, NJ, OR, LA, OH
Lot/Code Info: Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.