medical Class II Updated 2021-05-26

Bard Access Systems Inc. recalls REF 9617408D, Groshong NXT ClearVue Catheter with Sherlock T

Recalled Product

REF 9617408D, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Maximal Barrier Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036016 - Product Usage: is designed for use when central venous catheterization is prescribed.


Hazard / Issue

Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot No. reen4697 and reep0632
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →