medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge- BEQ-TOP 39202 NICU ECC PACK Material: 701

Recalled Product

Maquet Getinge- BEQ-TOP 39202 NICU ECC PACK Material: 701067313R01


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000043846 3000047255 3000052044 3000061508 3000074098 3000078928 3000079607