medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge- (1) Material: 701067350R01 BEQ-TOP 5208

Recalled Product

Maquet Getinge- (1) Material: 701067350R01 BEQ-TOP 5208 ECC Small-Cardiac (2) material: 701067350R02 BEQ-TOP 5208 ECC Small-Cardiac


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: (1) 3000067384, 3000070970 (2) 3000074734, 3000081003
View official government recall