medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge- (1) Material :701067370R01 BO-TOP 38803 OPE

Recalled Product

Maquet Getinge- (1) Material :701067370R01 BO-TOP 38803 OPEN HEART PACK W (2) Material: 701067370R02 BO-TOP 38803 OPEN HEART PACK W (3) Material: 701067370R03 BO-TOP 38803 OPEN HEART PACK W


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: (1) 3000042627 (2) 3000047034 (3) 3000058143
View official government recall