medical Class II Updated 2018-05-16

Siemens Healthcare Diagnostics, Inc. recalls Dimension¿ Sirolimus (SIRO) Flex¿ reagent cartridge

Recalled Product

Dimension¿ Sirolimus (SIRO) Flex¿ reagent cartridge


Hazard / Issue

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: SIRO, sirolimus test system, DF306, SMN# 10464331 Lot codes: BB7336 FA8037 FA8081 BB8167 FB8177 FB8219