medical Class II Updated 2018-05-16

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Mass Creatine Kinase (MMB) Flex¿ reagent ca

Recalled Product

Dimension Vista¿ Mass Creatine Kinase (MMB) Flex¿ reagent cartridge


Hazard / Issue

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: MMB, colorimetric method, cpk or isozymes, K6420, SMN# 10445097 Lot codes: 16284BD 16312BB 17011BD 17058BD 17093BC 17136BE 17186BE 17235BC 17268BA