medical Class II Updated 2020-04-15

Beckman Coulter Inc. recalls Beckman Coulter MicroScan PROMPT Inoculation System-D, Catal

Recalled Product

Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.


Hazard / Issue

Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot number 20201103T, Expiration Date: 11/3/2020
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →