Siemens Healthcare Diagnostics, Inc. recalls IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Anti HBc

Recalled Product

IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Anti HBc


Hazard / Issue

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Anti HBc, Anti HBc (IMMULITE 2000/IMMULITE 2000 XPi L2KHC2 (D)) Lot codes: Anti HBcM # 579 580 581 D579 D580 D581 ***Added 9/11/20*** D602