Belmont Instrument LLC recalls 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-0
Recalled Product
3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.
Hazard / Issue
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.