Radiometer Medical ApS recalls ABL800 FLEX, model number/UDI 393-800/0570 In Vitro testing
Recalled Product
ABL800 FLEX, model number/UDI 393-800/0570 In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) In vitro testing of samples of expired air for the parameters pO2 and pCO2 In vitro testing of pleura samples for the pH parameter.0693938004 and 393-801/05700693938011 - Product Usage:
Hazard / Issue
Analyzer's barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample resulting in delayed medical treatment