medical Class II Updated 2018-05-16

Siemens Healthcare Diagnostics, Inc. recalls IMMULITE ¿ /IMMULITE ¿ 1000 Vitamin B12

Recalled Product

IMMULITE ¿ /IMMULITE ¿ 1000 Vitamin B12


Hazard / Issue

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Vitamin B12, Vitamin B12 (IMMULITE/IMMULITE 1000, LKVB1) Lot codes: 0338 0339 0340 0341
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →