medical Class II Updated 2019-06-05

Exactech, Inc. recalls Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 60

Recalled Product

Opteform Allograft Disc, 90mm x 5mm, 22cc, Catalog Number 600-03-90


Hazard / Issue

The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.

Issued by

FDA

Affected States: OK, VA, FL
Lot/Code Info: UDI 10885862094100 Serial Numbers: T31977961, T31977962, T31977968
View official government recall