medical Class II Updated 2020-04-22

Diagnostica Stago, Inc. recalls STA - Deficient VIII, Product Code: 00725. STA - Deficient

Recalled Product

STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979


Hazard / Issue

A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.
View official government recall

Looking for safe alternatives?