medical Class II Updated 2020-04-22

Sysmex America, Inc. recalls The CF-70 instrument (product code: LXG; Regulation Number:

Recalled Product

The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.


Hazard / Issue

Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog No. AH699752.
View official government recall