Flowonix Medical Inc recalls Prometra Programmer (Grand Prime) Software version 2.00.29,
Recalled Product
Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
Hazard / Issue
The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.