medical Class II Updated 2020-04-22

Canon Medical System, USA, INC. recalls Canon Aquilion Lightning, Model TSX-036A

Recalled Product

Canon Aquilion Lightning, Model TSX-036A


Hazard / Issue

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Issued by

FDA

Affected States: AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, WV
Lot/Code Info: Canon Aquilion Lightning, TSX-036A - Serial numbers 4LC1892122, 4LC1892125, 4LC1892129, 4LC1882114, 4OC1912163, and 4OC18Y2149.
View official government recall