medical Class II Updated 2020-04-22

Canon Medical System, USA, INC. recalls Canon Aquilion One Vision, Model TSX-301C

Recalled Product

Canon Aquilion One Vision, Model TSX-301C


Hazard / Issue

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Issued by

FDA

Affected States: AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, WV
Lot/Code Info: Canon Aquilion ONE, Whole-Body X-Ray CT Scanner, Model TSX-301C - Serial number 3DC18Y2118.
View official government recall