medical Class II Updated 2020-04-29

Wilson-Cook Medical Inc. recalls HEMO-7 Hemospray Endoscopic Hemostat

Recalled Product

HEMO-7 Hemospray Endoscopic Hemostat


Hazard / Issue

Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Lot numbers manufactured 01/16/2017 - 01/15/2020 of both the following: HEMO-7 Order Number G56572 HEMO-10 Order Number G21049
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →