medical Class II Updated 2020-04-29

Rayner Intraocular Lenses Limited recalls RaySert PLUS RSP01. LOT K030. Expiry of 2024-01 - Product Us

Recalled Product

RaySert PLUS RSP01. LOT K030. Expiry of 2024-01 - Product Usage: The single use injectors (Model STWOl and Model RSPOl) are intended to be used to compress and insert into the capsular bag only those intraocular lenses that allow the use of these injectors in their approved labelling.


Hazard / Issue

RaySert PLUS RSP01 LOT K030 labeled with an expiry of 2024-01 contained an inner tray that was labeled with an expiry of 2015. In addition to having an expiry of 2015 the LOT number on the inner tray is identified as V001.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: LOT K030.