medical Class I Updated 2025-05-28

C.R. Bard Inc recalls Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, RE

Recalled Product

Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32"), 41" Length (104cm) with 8" (20.3cm) Balloon X-Ray Opaque Rubber


Hazard / Issue

Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.

Issued by

FDA

Affected States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Lot/Code Info: All Lots within expiration/UDI: (01)00801741076824