medical Class II Updated 2020-04-29

Radiometer Medical ApS recalls ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usa

Recalled Product

ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.


Hazard / Issue

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All software versions below 3.4 MR2
View official government recall