medical Class II Updated 2020-04-29

Acutus Medical Inc recalls Acutus Medical AcQGuide Mini Fixed-Curve Introducer with AcQ

Recalled Product

Acutus Medical AcQGuide Mini Fixed-Curve Introducer with AcQCros QX Integrated Dilator/Transseptal Needle, Ref: 112852 Sterile EO, RxOnly


Hazard / Issue

It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.

Issued by

FDA

Affected States: CA, SC, AZ
Lot/Code Info: Affected Lot Numbers/Expiration Date: 19101021 /October 2020 19103124 /October 2020 19121604 /December 2020