Becton Dickinson & Company recalls Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10
Recalled Product
Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294, 10015414, 10933805, 11171447, 11404930, 11522558, 11607704, 10013361T, 2110-0500, 2126-0500, 2200-0500, 2411-0500, 2419-0007, 2420-0007, 2420-0500, 2421-0500, 2426-0007, 2426-0500, 2429-0500, 2433-0007, 2477-0000, 2477-0007 2478-0000 C24116E Infusion sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient s artery or vein.
Hazard / Issue
An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.