medical MODERATE Updated 2017-04-19

Volcano Corporation recalls Refinity Rotational Intravascular Ultrasound (IVUS) Catheter

See all recalls from Volcano Corporation →

Recalled Product

Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.289; Catalog number 89800. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

Issued by

FDA

Distribution: Distribution to US and Japan
Lot/Code Info: Lot # 50068951, Expiry date 06/30/2018; Lot # 50068952, Expiry date 06/30/2018; Lot # 50069133, Expiry date 06/30/2018; Lot # 50083387, Expiry Date 11/30/2018; Lot # 50083390, Expiry Date 11/30/2018.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall