medical Class II Updated 2018-05-16

Siemens Medical Solutions USA, Inc recalls Syngo.Via VB20A model 1049610

Recalled Product

Syngo.Via VB20A model 1049610


Hazard / Issue

Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All
View official government recall