medical Class I Updated 2023-06-28

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; I

Recalled Product

ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators


Hazard / Issue

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN 00643169720466, Lot Serial Numbers: BWI218211H, BWI218218H, BWI218226H, BWI218228H, BWI218206H, BWI218224H, BWI218210H, BWI218222H, BWI218217H, BWI218214H, BWI218216H, BWI218208H, BWI218207H, BWI218221H, BWI218212H, BWI218220H
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →