medical Class II Updated 2025-05-28

Echopixel, Inc. recalls TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software v

Recalled Product

TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system.


Hazard / Issue

Imaging system data may be displayed with an incorrect orientation (sidedness) where the patient side is not correct (left, right, superior, inferior, anterior or posterior). This may lead to an inaccurate measurements.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number EP-00001-XX with software versions 1.0, 1.25, and 1.35 UDI-DI Codes: 00851325007003 Serial Numbers: Field Products: C2015-09-0001 C2015-09-0003 D2015-05-0002 C2015-06-0002 D2015-12-0004 C2015-08-0001 D2015-12-0005 C2015-09-0002 D2015-12-0001 C2016-01-0001 D2015-12-0002 C2015-12-0002 D2015-12-0003 D2016-01-0002
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →