medical Class II Updated 2021-06-16

RAYSEARCH LABORATORIES AB recalls RayStation (Treatment Planning System) Model Number RayStat

Recalled Product

RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .


Hazard / Issue

Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software version: RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 2 including some service packs.