medical Class II Updated 2019-06-26

CooperSurgical, Inc. recalls LEEP Precision Integrated System 120V, Model LP-10-120 Pr

Recalled Product

LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .


Hazard / Issue

2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014, 2018110015,2018110016, 2018110017, 2018110019, 2018110020,2018110021,& 2018110022