medical Class II Updated 2018-05-23

Kelyniam Global, Inc. recalls Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003

Recalled Product

Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003


Hazard / Issue

Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C. Case Numbers: CSI111715-KH1 CSI072814-JB1 CSI082514-MA1 CSI080714-CT1 CSI091914-JW1 CSI101614-TR1 CSI041415-MZ1 CSI100915-DD1 CSI120215-MM1 CSI100915-DD2 CSI011816-TA1 CSI012016-RT1A CSI012716-MP1 CSI021116-CA1 CSI030716-BC1 CSI040716-CK1 CSI060916-DF1 CSI070716-AC1 CSI092116-JA1 CSI110416-DW1 CSI120616-LS1 CSI012417-SD1 CSI062617-HR1
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →