medical Class II Updated 2019-07-03

Thermo Fisher Scientific (Asheville) LLC recalls Thermo Scientific General Purpose Dynamic Bath Precision 2

Recalled Product

Thermo Scientific General Purpose Dynamic Bath Precision 28L GP Bath Model Number: TSGP28


Hazard / Issue

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Issued by

FDA

Affected States: CA, GA, KY, PA
Lot/Code Info: Serial Numbers: 300246819 300247423 300247424