medical Class III Updated 2019-07-03

GE Healthcare, LLC recalls Replacement leadwires with grabber ends: REF/Catalog Number

Recalled Product

Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.


Hazard / Issue

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: product does not have lot or serial codes No GTIN/UDI
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →