medical Class II Updated 2020-05-13

Intuitive Surgical, Inc. recalls daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, R

Recalled Product

daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.


Hazard / Issue

The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot numbers: T10180511; T10180718; T10180821;T10180822; T10180906; T10180913; T10180925; T10181008; T10190204; T10190208; T10190213; T11180816; T11190219