Siemens Medical Solutions USA, Inc recalls Siemens Artis Q floor --Interventional Fluoroscopic X-Ray Sy

Recalled Product

Siemens Artis Q floor --Interventional Fluoroscopic X-Ray System Model Number: 10848280


Hazard / Issue

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 103008 103030 103028 103258 103221 103222 103277 103220 103279 103285 103286 103295