medical Class II Updated 2020-05-13

Delfi Medical Innovations, Inc. recalls Delfi Medical Innovations, VariFit Contour Cuffs, single use

Recalled Product

Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile


Hazard / Issue

Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.

Issued by

FDA

Distribution: US: Ohio OUS: Austrailia
Lot/Code Info: 18J17
View official government recall