medical Class II Updated 2021-06-30

Philips Medical Systems Nederlands recalls Philips Allura Xper System: Interventional Fluoroscopic X-Ra

Recalled Product

Philips Allura Xper System: Interventional Fluoroscopic X-Ray System Product No.: 722026, 722027, 722028, 722029, 722035, 722038, and 722058.


Hazard / Issue

Certain low-voltage DC power supplies (DCPS) used in these products may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.

Issued by

FDA

Affected States: AL, AZ, FL, GA, IL, MA, MO, NJ, OR, VA
Lot/Code Info: Product No.: 722026, 722027, 722028, 722029, 722035, 722038, and 722058. Date(s) of Distribution: 16-SEP-2014 - 04-MAY-2016. System equipment numbers: 64937259 65862117 63304032 63151474 63250793 63303889 63304149 64140063 63855729 63516113 64077887 66407452 66350303 65342973 63210187 63511276; 66923399 66422060 64085323 61099178 64103799 67936786 66730784 66736708 66628236 63883590 64916638 64927462 65345691 66517161 66403768 63218881 67360962 69856976 65258983 66167983 64778594 61655917 65206981 67874859