medical HIGH RISK Updated 2024-06-19

Thoratec Corp. recalls Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A,

See all recalls from Thoratec Corp. →

Recalled Product

Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00813024010142, 00813024010852. All Serial Numbers

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.