medical MODERATE Updated 2026-04-29

Remel Europe Ltd. recalls Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

See all recalls from Remel Europe Ltd. →

Recalled Product

Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858