Philips North America, LLC recalls Philips Zenition 70, System code 718133 - Product Usage: The

Recalled Product

Philips Zenition 70, System code 718133 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.


Hazard / Issue

When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers 35 37 56 20 50 2 7 8 10 12 13 47 67 34 19 31 45 86 30 59 51 21 54