medical Class II Updated 2018-06-06

Medline Industries Inc recalls MEDLINE O.R. Necessities(TM) Sterile* Needle Counter, 100 Ct

Recalled Product

MEDLINE O.R. Necessities(TM) Sterile* Needle Counter, 100 Ct. Double Magnet / Foam Block with Blade Remover, reorder: NC100FBR Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.


Hazard / Issue

The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 67016114502 67016124501 67016124502 67016124507 67017014501 67017024502 67017034501 67017034503 67017064502 67017074502 67017084502 67017084503 67017094501 67017094503 67017094505 67017104501 67017114501 67017114502
View official government recall

Looking for safe alternatives?