medical Class II Updated 2020-05-20

Beckman Coulter Inc. recalls iChem¿VELOCITY Urine Chemistry System-

Recalled Product

iChem¿VELOCITY Urine Chemistry System-


Hazard / Issue

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog #700-7176-001 and #700-7177-001, All part numbers
View official government recall