medical HIGH RISK Updated 2025-06-25

Dexcom, Inc. recalls Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011

See all recalls from Dexcom, Inc. →

Recalled Product

Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Lots/ UDI: STK-AT-011 - 00386270002198 STK-AT-012 -00386270002518 STK-AT-013 - 00386270002525 STK-GT-001 - 00386270004291 STK-GT-008 - 00386270004260 STK-GT-013 - 00386270002273 STK-GT-019 - 04595986293080 STK-GT-100 - 00386270002020 STK-GT-109 - 00386270002129 STK-GT-113 - 00386270003638

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Dexcom G7 Glucose Receiver, →
View official government recall