medical Class II Updated 2020-05-20

Beckman Coulter Inc. recalls iQ200 Series Urine Microscopy Analyzer- All part numbers

Recalled Product

iQ200 Series Urine Microscopy Analyzer- All part numbers


Hazard / Issue

A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Numbers: 00-3322, 700-3325, 700-3326, 700-3345, 700-3347, 700-3370, 700-3375, 700-3300RFB, B48999, C10683, C10684., All part numbers
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →