medical Class II Updated 2019-07-17

Rinovum Women's Health recalls Revive Reusable Bladder Support, Model No. 0025 - Product Us

Recalled Product

Revive Reusable Bladder Support, Model No. 0025 - Product Usage: The Revive bladder support is intended for temporary management of urinary leakage caused by stress urinary incontinence (SUI) in women, 18 years and older. The device comes in a sealed box, and contains a reusable bladder support and applicator, and disposable strings and travel case. The device is nonperishable and does not have an expiration date.


Hazard / Issue

The product is unable to meet the use life of 31 cycles and may split or crack at the base of the "petals" after repeated use but within the use life of the product (31 uses).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot 75770223
View official government recall