medical Class II Updated 2020-05-20

Siemens Medical Solutions USA, Inc recalls Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11

Recalled Product

Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.


Hazard / Issue

Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 40057 40047 40048 40107 40108 40111 40120 40126 40134 40135 40032 40104 40075 40102 40124 40070
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →