medical Class II Updated 2023-07-05

Ellex Medical Pty Ltd. recalls Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre

Recalled Product

Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G


Hazard / Issue

When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Nos., UDI IPY0374, (01) 0 9342395 00012 0 ISRY0242, (01) 0 9342395 00015 1 ISG0370, (01) 0 9342395 00024 3 IPG0539, (01) 0 9342395 00009 0 BG1190