medical Class II Updated 2021-07-07

Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Op

Recalled Product

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/728201


Hazard / Issue

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model/ Serial Number/UDI: 728120 2516 N/A 728120 3179 N/A 728120 3316 N/A 728201 9326 N/A 728120 3304 N/A 728120 3271 N/A 728120 2537 N/A 728120 3324 N/A 728120 H3321 N/A 728120 3058 N/A 728125 2529 N/A